JSESD005: Development and Implementation of a Low-cost Source to-surface Distance (SSD), Collimator Rotation and Gantry Angle Verification Device and its Implications in the Medical Field

Authors

  • Prince Chitopho Author
  • Innocent Muchingami Author
  • Gilbert Mawera Author

Keywords:

collimator rotation measurement, gantry angle, source-to-surface distance (SSD), QC verification device

Abstract

The demand for radiotherapy services in developing countries cannot 
be overemphasised; an increase in cancer cases has been observed and 
governments have invested heavily in cancer treatment through procurement, 
maintenance and service of radiotherapy equipment and accessories. 
The availability of such equipment facilitates the management of various 
cancer cases by offering services ranging from diagnosis to palliative and 
cure. However, quality control devices would be required for the treatment 
machines to operate according to specified standards. These verification 
devices are so expensive that most developing countries can only buy entry 
point devices. The basic verification devices in most hospitals are mechanical 
apparatus like the magnetic front pointer (for source to source-to-surface 
distance) and spirit level for gantry angle verification. In this study, a linear 
accelerator (Linac) gantry collimator angle and source-to-surface distance 
(SSD) verification device was developed and implemented. The physical 
parameters (distance and angle) are measured using sensor modules. The 
Atmega 328 microcontroller is used as the main controller for this device. 
Low-cost and locally-available materials are used in the development of 
the device. The device can also measure time and temperature, making it 
a four-in-one multi-purpose QC verification device. The device was tested 
and calibrated with reference to secondary standards in a local government 
hospital setup. A linear relationship was observed between the device and 
equipment readouts, and the device was then characterised to match the 
actual measurements. To facilitate inferences, a comparison of results was 
done using the appropriate statistical tools. Correlation of 99.9996% and 
a confidence level of 99.98% were observed between the data sets. Through 
the use of the device, about 81% and 19% was observed in time savings for 
distance and angle measurement, respectively. The usage of this device brings 
various operational features such as accuracy, ease of use and increased 
productivity. 