JSESD005: Development and Implementation of a Low-cost Source to-surface Distance (SSD), Collimator Rotation and Gantry Angle Verification Device and its Implications in the Medical Field
Keywords:
collimator rotation measurement, gantry angle, source-to-surface distance (SSD), QC verification deviceAbstract
The demand for radiotherapy services in developing countries cannot
be overemphasised; an increase in cancer cases has been observed and
governments have invested heavily in cancer treatment through procurement,
maintenance and service of radiotherapy equipment and accessories.
The availability of such equipment facilitates the management of various
cancer cases by offering services ranging from diagnosis to palliative and
cure. However, quality control devices would be required for the treatment
machines to operate according to specified standards. These verification
devices are so expensive that most developing countries can only buy entry
point devices. The basic verification devices in most hospitals are mechanical
apparatus like the magnetic front pointer (for source to source-to-surface
distance) and spirit level for gantry angle verification. In this study, a linear
accelerator (Linac) gantry collimator angle and source-to-surface distance
(SSD) verification device was developed and implemented. The physical
parameters (distance and angle) are measured using sensor modules. The
Atmega 328 microcontroller is used as the main controller for this device.
Low-cost and locally-available materials are used in the development of
the device. The device can also measure time and temperature, making it
a four-in-one multi-purpose QC verification device. The device was tested
and calibrated with reference to secondary standards in a local government
hospital setup. A linear relationship was observed between the device and
equipment readouts, and the device was then characterised to match the
actual measurements. To facilitate inferences, a comparison of results was
done using the appropriate statistical tools. Correlation of 99.9996% and
a confidence level of 99.98% were observed between the data sets. Through
the use of the device, about 81% and 19% was observed in time savings for
distance and angle measurement, respectively. The usage of this device brings
various operational features such as accuracy, ease of use and increased
productivity.